Open original FDA record ↗Recall record
- FDA identifier
- Z-2227-2018
- Company in FDA record
- The Seaberg Company Inc
- FDA record creation date
- Jun 14, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Based on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial blood flow. There is a remote chance that a serious injury and/or death could occur as a result of the failure of the device.
Source facts
- Action
- On 05/02/2018, the firm posted a press release notifying customers of a recall of all unused SAM XT Extremity Tourniquets (SAM XT) due to a possible failure of the stitches securing the buckle to the nylon belt that could pose a potential risk when used on a human patient to stop arterial blood flow. On 05/03/2018, recall notification forms were mailed to customers via overnight mail service.
- Initiated
- 2018-05-02
- Root cause
- Employee error
- Terminated
- 2020-07-01
- Distribution
- USA: CA, MD, OH, FL, ID, CO, CT, OR, AZ, IL, ME, MO, NY, SC, MI, AR, MT, CT, WA, TX, KY, NC, NJ, PA OUS: Australia, Canada, China, Columbia, Costa Rica, Europe (EU), Hong Kong, Iceland, India, Iraq, Israel, Japan, Kenya, Korea, Lebanon, Malaysia, Russia, Saudi Arabia, Singapore, South Africa, Thailand, UAE
- Recall number
- Z-2227-2018
- Recall status
- Terminated
- CDRH record ID
- 164667
- FDA record created
- 2018-06-14
- Recall event number
- 80140