Open original FDA record ↗Recall record
- FDA identifier
- Z-2227-2021
- Company in FDA record
- Carl Zeiss Meditec, Inc.
- FDA record creation date
- Aug 6, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to failure to acquire pre-market clearance for its high resolution tomographic/biomicroscope imaging device (Model 6000) prior to commercializing the device.
Source facts
- Action
- On 07/08/2021 the firm sent an "Medical Device Recall" Notification Letter via UPS informing customers that it commercialized in the U.S. a device without receiving a substantial equivalence clearance [510(k)]. The Recalling Firm is informing and instructing its customers, that while it is working on submitting and received a 510(k) clearance, the affected system may continued to be used as normal. The Recalling Firm has placed a ship hold on all distribution for the U.S. market, but will continue to provide product support for current customers. Customers are also instructed to complete and return the Acknowledgement and Receipt Form via email at USFCA.meditec.us@zeiss.com or via Fax at 925-448-3880. For any questions, contact Customer Service at 877-486-7473.
- Initiated
- 2021-07-08
- Root cause
- No Marketing Application
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY O.U.S.: N/A
- Recall number
- Z-2227-2021
- Recall status
- Open, Classified
- CDRH record ID
- 187906
- FDA record created
- 2021-08-06
- Recall event number
- 88109