Z-2228-2026: Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (72

Recall record Z-2228-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2228-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
May 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

Source facts

Action
On April 28, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case the affected system has been transferred to another organization, please send a copy of this URGENT Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 4. If you experience the issue described in this letter, please report it to your local Philips representative. Actions planned by Philips IGT Systems to correct the issue Starting in Q1 2027, Philips will replace the deaeration hose in the CU3101 tube cooling unit in all affected systems. Your local Philips representative will contact you to schedule a visit to replace the deaeration hose once available.
Initiated
2026-04-28
Root cause
Component design/selection
Distribution
Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;
Recall number
Z-2228-2026
Recall status
Open, Classified
CDRH record ID
219852
FDA classified
2026-05-21
Recall event number
98801

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗