Open original FDA record ↗Recall record
- FDA identifier
- Z-2229-2009
- Company in FDA record
- St. Jude Medical
- FDA classification date
- Sep 18, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
St. Jude Medical has initiated a voluntary recall of two lots of Fast-Cath Hemostasis Introducers-reorder number 406709, lot number 4219 and reorder number 406702, lot number 6372. They discovered that the product lots could contain some items which have been mis-labeled with the incorrect French size. In one case, the pouch label does not correctly identify the content of the pouch. In another case, the carton label is incorrect. The pouch label indicated that the product was 4F catheter, but the contents of the pouch may be a 7F catheter. St. Jude Medical has not received any reports of injuries resulting from this mis-labeling.
Source facts
- Action
- Field Representatives were sent a "St Jude Medical Urgent Product Information" letter to be hand delivered to their customers. The letter listed the products and problem and advised consignees to discontinue use of the product. A form was attached for the return of products to the firm. Contact your local St Jude Medical representative for further information.
- Initiated
- 2009-08-17
- Root cause
- Error in labeling
- Terminated
- 2011-12-17
- Distribution
- Worldwide Distribution -- US (CA, MI, PA and TN), China and Belgium.
- Recall number
- Z-2229-2009
- Recall status
- Terminated
- CDRH record ID
- 84689
- FDA classified
- 2009-09-18
- Recall event number
- 53121