Open original FDA record ↗Recall record
- FDA identifier
- Z-2230-2009
- Company in FDA record
- ArjoHuntleigh
- FDA classification date
- Sep 22, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A Scoop Stretcher is made up of two frames that are held together with two locking devices, one on each end. Each locking device contains a safety latch. The safety latch could unintentionally release when the side frames are pulled for positioning.
Source facts
- Action
- ArjoHuntleigh sent their customers a cover letter and the Field Safety Notification, informing them of the potential problem with the safety latches on the stretcher and a pending visit from an ArjoHuntleigh Service Technician to correct the problem. These letters were sent via UPS 2nd day delivery on 8/18/09. The customers were requested to test their stretchers upon receipt of the letter following the instructions for checking the safety latches on the stretcher: * Place the stretcher on its side on a non-scratch surface. * Try to pull the two sides apart with the safety latch engaged. * Test both the head and foot ends of the stretcher. * If the stretcher pulls apart, remove it from use immediately. It is important to note that all affected stretchers will be upgraded, whether they fail the test or not.
- Initiated
- 2009-08-18
- Root cause
- Manufacturing material removal
- Terminated
- 2010-01-07
- Distribution
- Worldwide Distribution
- Recall number
- Z-2230-2009
- Recall status
- Terminated
- CDRH record ID
- 84080
- FDA classified
- 2009-09-22
- Recall event number
- 52824