Z-2230-2019: iGUIDE System with software build iGuide 2.2.0, 2.2.1, and 2.2.2

Recall record Z-2230-2019 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2230-2019
Company in FDA record
Elekta, Inc.
FDA record creation date
Aug 9, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for iGUIDE to incorrectly monitor the 3D position.

Source facts

Action
Elekta, Inc. notified customers on about 06/19/2019 via URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter. Customers were instructed to follow the below steps: 1. Log on to iGUIDE and move the 3D position. If the 3D position is equal to the PARK position, the system is not affected. 2. If the 3D position is not equal to the PARK position, log on to iGUIDE with Operator and Administrator rights. (If no user is available, go to the iGUIDE User Manager and add an appropriate user.). 3. Move the HexaPOD to the 3D Position (If 3D position is equal to the DRIVE position, both checkmarks will appear). 4. Open the HexaPOD properties (Left click the HexaPOD symbol on the system overview), which will display the HexaPOD properties. 5. Enter a rotation of 1.00 degree for any of the axis (RX, RY, RZ) and click Go to position. 6. Execute the pending movement. If the checkmark for the 3D position disappears and the EXTERNAL INHIBIT is set, your system is not affected by the comma / dot issue. No further actions are necessary. If the 3D position monitoring does not consider the 1.00 degree and the checkmark for the 3D position is still displayed and the EXTERNAL INHIBIT is not set, your system is affected! If the customer wants to still use a 3D workflow without imaging, do not rely on the 3D position indicator in iGUIDE. The HexaPOD should be moved to the PARK position first, then the 3D workflow can be started moving the HexaPOD back to the 3d position. Customers were also instructed to place the recall notice in a place accessible to all users (e.g. Instructions for Use, until this action is closed) and to advise the appropriate personnel, working with this product, on the content of this letter. A technical solution (iGUIDE patch) will be provided to correct the behavior. Until the iGUIDE patch is available, an interim solution can be implemented (change comma to dot) if your system is affected. Please consult your local Elekta represent
Initiated
2019-06-19
Root cause
Software Design Change
Terminated
2024-05-06
Distribution
Nationwide distribution to AK, AL, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI. Worldwide distribution to Algeria, Argentina, Australia, Austria, Belgium, Botswana, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Egypt, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.
Recall number
Z-2230-2019
Recall status
Terminated
CDRH record ID
174163
FDA record created
2019-08-09
Recall event number
83180

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗