Reason stated in the source
Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample.
Source facts
- Action
- Firm notified all ARCHITECT c8000 customers with a correction letter and response form distributed on 5/23/2008. The letter provided actions the users could take until an updated software version was available. For assistance, contact Abbott at 1-877-4ABBOTT.
- Initiated
- 2008-05-23
- Root cause
- Software design
- Terminated
- 2010-04-26
- Distribution
- Worldwide distribution, including: USA (including Puerto Rico), Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, France, Germany, Grand Cayman, Greece, Guatemala, Honduras, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Uruguay, and Venezuela.