Z-2231-2008: ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected s

Recall record Z-2231-2008 for Abbott Laboratories, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2231-2008
Company in FDA record
Abbott Laboratories, Inc
FDA classification date
Sep 11, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample.

Source facts

Action
Firm notified all ARCHITECT c8000 customers with a correction letter and response form distributed on 5/23/2008. The letter provided actions the users could take until an updated software version was available. For assistance, contact Abbott at 1-877-4ABBOTT.
Initiated
2008-05-23
Root cause
Software design
Terminated
2010-04-26
Distribution
Worldwide distribution, including: USA (including Puerto Rico), Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, France, Germany, Grand Cayman, Greece, Guatemala, Honduras, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Uruguay, and Venezuela.
Recall number
Z-2231-2008
Recall status
Terminated
CDRH record ID
71410
FDA classified
2008-09-11
Recall event number
48431

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗