Z-2231-2025: 3M Ranger Irrigation Fluid Warming Set, REF 24750

Recall record Z-2231-2025 for 3M Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2231-2025
Company in FDA record
3M Company
FDA classification date
Aug 1, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Notice was issued to clarify flow rates related to the inlet fluid temperature.

Source facts

Action
Solventum issued an Urgent Medical Device Correction notice dated 04/21/2025 to its consignees via email and US certified mail. The notice explained the issue, the modification related to the inlet fluid temperature to address the issue, and forthcoming instructions for use updates. All users of the impacted products are being asked to take the following actions. " Ensure this notification is passed on to all those who need to be aware within your organization and to any department where the affected products have been transferred. " Continue to use the products as indicated, considering the updated information provided. " Keep this notice on hand until you are notified that the updated IFUs and operator s manuals are available. " Complete the attached Customer Acknowledgement Form acknowledging that you have read and understood this letter and the updated information provided. " Email the completed form to 3MFieldSafetyAction@montage72.com. For additional questions or need to report an adverse event, contact the Solventum [formerly 3M Health Care] Customer Helpline at 1-800-228-3957, option 6.
Initiated
2025-04-22
Root cause
Device Design
Distribution
US (nationwide) and Argentina, Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, Canary Islands, China, Costa Rica, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Korea, Luxembourg, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-2231-2025
Recall status
Open, Classified
CDRH record ID
214727
FDA classified
2025-08-01
Recall event number
97174

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗