Skip to content

Public FDA record

Z-2232-2008: Hamilton Soft Grip Pipettes, 1 ml fixed volume, model number 55019-19 Product is indicated for manual pip

Recall record Z-2232-2008 for Hamilton Co. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2232-2008
Company in FDA record
Hamilton Co
FDA classification date
Sep 17, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacity

Source facts

Action
Recall initiated May 7 2008. Notification issued via certified overnight mail explains the problem and instructs customers to inventory the affected pipettes. The firm will contact its customers and supply replacement lower-body assemblies, repair instructions and return information so that affected parts can be returned to Hamilton. A fax-back form is supplied with each notification. Responses will be tracked. Contact Hamilton Co. at 1-775-558-3000 for assistance.
Initiated
2008-05-07
Root cause
Device Design
Terminated
2011-02-08
Distribution
Product distributed throughout the US and to Puerto Rico, New Zealand, Australia, Japan and Canada.
Recall number
Z-2232-2008
Recall status
Terminated
CDRH record ID
71359
FDA classified
2008-09-17
Recall event number
48425

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗