1) The monitor's ability to withstand electrostatic discharge (ESD) to exposed metal connectors on the back of the monitor which may cause the monitor's display to freeze.
2) The unit generates electromagnetic radiation above the limits set by applicable standards for medical devices
Source facts
Action
Oridian sent a letter to all its distributors June 25, 2009, instructing them to gather all the units to be exchanged for factory reconditioned units that have the hardware and software enhancements and meet all performance requirements in full.
Initiated
2009-06-30
Root cause
Device Design
Terminated
2010-10-01
Distribution
Nationwide.
Recall number
Z-2232-2009
Recall status
Terminated
CDRH record ID
84332
FDA classified
2009-09-21
Recall event number
52959