Z-2233-2024: EasySuite 4K Integrated Operating Room System Model/Catalog Number: EasySuite 4K- The EasySuite 4K system

Recall record Z-2233-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2233-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Jun 28, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A subset of the ES4K systems do not have complete records for earth leakage testing. The earth leakage test is a series of measurements intended to confirm that the amount of electrical current flowing to the earth is within standard limits

Source facts

Action
Olympus issued Urgent: Medical Device Corrective Action letter on May 24, 2024 via Courier, 2-day delivery. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to schedule an on-site service to perform the earth leakage testing on your facility s ES4K system. You may continue to use your ES4K system until an Olympus representative performs the earth leakage test. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. If you have further distributed this product, identify your customers, and forward them this notification. 3. Acknowledge receipt of this letter through our Olympus recall portal: a. Go to https://olympusamerica.com/recall. b. Enter the recall number "0446 c. Complete the form as instructed and include your account ID number. 4. An Olympus representative will contact you to complete the earth leakage test which will serve as the completion of this field corrective action for your facility. Olympus requests that you report any complaints, including surgical displays becoming inoperable/unusable or touch current sensation (i.e., vibration, slight tingle), to the Technical Assistance Center (TAC) at 1- 800-848-9024, option 1.
Initiated
2024-04-24
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Hong Kong.
Recall number
Z-2233-2024
Recall status
Open, Classified
CDRH record ID
208094
FDA classified
2024-06-28
Recall event number
94711

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗