Skip to content

Public FDA record

Z-2234-2009: S&W Electrodes for Defibrillation, 10 Defibrillation Electrodes, Reorder #90101, S & W Healthcare Corp, 1

Recall record Z-2234-2009 for Leonhard Lang Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2234-2009
Company in FDA record
Leonhard Lang Gmbh
FDA classification date
Sep 24, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A potential risk was identified associated with the connectors of the defibrillation electrodes DF20. It was found that some of the connectors do not slip easily into the sockets of the defibrillator or an extension cord of the defibrillator. In such cases increased force or some wiggling is required to insert the connector fully. A potential risk for patients exist, if (a) such a connector (b)

Source facts

Action
A Leonhard Lang letter dated June 9, 2009 was sent June 6, 2009 to its customer. The letter identified the problem and the product. The letter instructed the consignee to stop using or distributing the product and to quarantine them. It also requested them to report all quantities of product that they had in stock using the attached form.
Initiated
2009-07-10
Root cause
Nonconforming Material/Component
Terminated
2011-12-17
Distribution
Worldwide Distribution - US - FL Foreign: Austria, Germany, Romania, France, Dubai, Greece, Italy, South Korea, United Kingdom, Hungary, Latvia, Saudi Arabia and Lebanon
Recall number
Z-2234-2009
Recall status
Terminated
CDRH record ID
84608
FDA classified
2009-09-24
Recall event number
52794

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗