Open original FDA record ↗Recall record
- FDA identifier
- Z-2234-2009
- Company in FDA record
- Leonhard Lang Gmbh
- FDA classification date
- Sep 24, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A potential risk was identified associated with the connectors of the defibrillation electrodes DF20. It was found that some of the connectors do not slip easily into the sockets of the defibrillator or an extension cord of the defibrillator. In such cases increased force or some wiggling is required to insert the connector fully. A potential risk for patients exist, if (a) such a connector (b)
Source facts
- Action
- A Leonhard Lang letter dated June 9, 2009 was sent June 6, 2009 to its customer. The letter identified the problem and the product. The letter instructed the consignee to stop using or distributing the product and to quarantine them. It also requested them to report all quantities of product that they had in stock using the attached form.
- Initiated
- 2009-07-10
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-12-17
- Distribution
- Worldwide Distribution - US - FL Foreign: Austria, Germany, Romania, France, Dubai, Greece, Italy, South Korea, United Kingdom, Hungary, Latvia, Saudi Arabia and Lebanon
- Recall number
- Z-2234-2009
- Recall status
- Terminated
- CDRH record ID
- 84608
- FDA classified
- 2009-09-24
- Recall event number
- 52794