Open original FDA record ↗Recall record
- FDA identifier
- Z-2234-2010
- Company in FDA record
- Hospira Inc.
- FDA classification date
- Aug 17, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Power Pack Accessory is not certified to UL 60601 standard or Subject 1363A for relocatable power taps intended for use with medical equipment. Potential problems could include electrical shock, electrical short or electrical fire.
Source facts
- Action
- Hospira sent Urgent Device Recall letters dated 7/1/10 to their direct accounts on 7/2/10 via Federal Express 2 day delivery, informing them of the potential for electrical shock, electrical short, or electrical fire, since the power packs were not certified to UL 60601 standard or Subject 1363A of UL standard 1363 for relocatable power taps intended for use with medical equipment. The accounts were instructed to examine their inventory for the affected product and quarantine it immediately, and Stericycle at 1-800-321-7388 to obtain a return kit for returning the affected product. If they distributed the affected product further, they were instructed to sub-recall the product from their sub-accounts. They were also requested to complete the enclosed reply form and return it to Hospira by fax at 1-262-577-6921, indicating the amount of product placed on hold for return, if they further distributed the product and if they notified their retail customers. For medical inquiries, the accounts were directed to contact Hospira Medical Communications at 1-800-615-0187.
- Initiated
- 2010-07-02
- Root cause
- Device Design
- Terminated
- 2011-07-20
- Distribution
- Worldwide Distribution -- US and Canada.
- Recall number
- Z-2234-2010
- Recall status
- Terminated
- CDRH record ID
- 92726
- FDA classified
- 2010-08-17
- Recall event number
- 56138