Analysis showed that the screwdriver was not fully inserted in the setscrew's hex cavity, which damaged the setscrew.
Source facts
Action
A Reply Technical Note, dated September 2008, titled Additional Instructions for use at Implant for REPLY pacemakers, was issued to physicians on September 24, 2008. The Technical Note describes the background for an increased incidence of complaints relating to the screwdriver and setscrew. The letter provided a reminder for inserting the screwdriver, and additional instructions for insertion of the screwdriver.
Initiated
2008-09-23
Root cause
Employee error
Terminated
2011-12-17
Distribution
Nationwide Distribution -- USA, including states of AL, AZ, CA, CO, CT, FL, GA, KS, KY, LA, MA, MI, NC, NJ, OH, OK, SC, TN, and TX.
Recall number
Z-2235-2009
Recall status
Terminated
CDRH record ID
75453
FDA classified
2009-09-24
Recall event number
50473