Open original FDA record ↗Recall record
- FDA identifier
- Z-2236-2010
- Company in FDA record
- Olympus America Inc.
- FDA classification date
- Aug 18, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An investigation of consumer complaints revealed the inability of the snare wire to be retracted into the snare tube after deployment of the snare.
Source facts
- Action
- Firm initiated this recall following investigation of customer complaints regarding the inability of the snare wire to be retracted into the snare tube. Customers were notified via certified mail, an Urgent: Medical: Device Recall letter issued on May 4, 2010. The letter identified the affected product and the reason for the recall. Customers were instructed to inspect their inventory and identify the affected product. Customers are to remove from use any identified inventory, quarantine, and discard them. Customers should contact the firm to obtain replacement products. Questions should be directed towards Laura Storms-Tyler at 484-896-5688 or laura.storms-tyler@olympus.com.
- Initiated
- 2010-05-03
- Root cause
- Other
- Terminated
- 2013-05-20
- Distribution
- Nationwide Distribution -- AR, CA, CO, CT, DC, FL, GA, IN, KS, LA, MD, MO, NC, NY, OH, PA, SC, TN, and TX.
- Recall number
- Z-2236-2010
- Recall status
- Terminated
- CDRH record ID
- 91483
- FDA classified
- 2010-08-18
- Recall event number
- 55663