Open original FDA record ↗Recall record
- FDA identifier
- Z-2236-2014
- Company in FDA record
- Nanosphere, Inc.
- FDA classification date
- Aug 14, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Nanosphere Inc. has recently determined through four customer complaints that a small number of Verigene Gram-Negative Blood Culture Nucleic Acid Test (BC-GN) Extraction Trays in Lot 042914021B have low levels or are missing Magnetic Microparticles (MMPs) required for DNA Extraction.
Source facts
- Action
- Nanosphere sent an Urgent Product Recall letter dated July 9, 2014 to all affected customers. All customers were initially notified of the recall by telephone on July 9, 2014. The customers were instructed to quarantine and return all affected product to Nanosphere. The letter included instructions: 1) on how to return the recalled products, and 2) to review any test results that were generated while using the recalled products to ensure that there has been no negative impact on patient care. Customers with questions were instructed to contact Nanosphere at 888-837-4436, option 2 or via e-mail at Technicalsupport@nanosphere.us.
- Initiated
- 2014-07-09
- Root cause
- Under Investigation by firm
- Terminated
- 2015-09-01
- Distribution
- Nationwide Distribution including CA, CO, IA, MT, OH, OK, OR, TX, and UT
- Recall number
- Z-2236-2014
- Recall status
- Terminated
- CDRH record ID
- 128801
- FDA classified
- 2014-08-14
- Recall event number
- 68831