Open original FDA record ↗Recall record
- FDA identifier
- Z-2236-2016
- Company in FDA record
- GE Inspection Technologies, LP
- FDA classification date
- Jul 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Radiation emissions which exceeded the limit in the performance standard were detected during maintenance of one system. The identified cause was insufficient attachment of shielding in the door. Jostling during delivery was identified as the probable cause of this failure to comply with the performance standard.
Source facts
- Action
- GE plans to notify its customers with the systems through a TIL. The TIL will inform customers that GE will add lead strips to the inside window frame under the lead glass window of the main cabinet door. GE recommends that it visit each customer site and repair the system by adding lead strips along the window frame of the cabinet door, beneath the lead glass window, within 90 days following FDA's approval of the repair plan.
- Initiated
- 2016-03-02
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-08-07
- Distribution
- US Distribution to the states of : FL, TN, and TX.
- Recall number
- Z-2236-2016
- Recall status
- Terminated
- CDRH record ID
- 147897
- FDA classified
- 2016-07-23
- Recall event number
- 74554