Z-2236-2021: Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic su

Recall record Z-2236-2021 for NEOSTEO. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2236-2021
Company in FDA record
NEOSTEO
FDA record creation date
Aug 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Traces of corrosion on the laser marking was detected on cannulated graduated drill bit, which could lead to the emission of wear / corrosion particles in the operated area, potentially leading to the following risks: Inflammation, pain, allergy, adverse tissue reaction. Device description: The drill bit is used to create the hole required for the introduction of the screw, it is suitable for Neosteo compression screws for cortical and spongy bone. Drilling depth can be read on the instrument. The shaft is coupled with the 4.5mm surgical motor using a AO 4.5mm male drive connexion. The countersink bit is coupled to a 1.0mm k-wire to ensure guiding while drilling. The instrument is packaged in a plastic sheath sealed by welding.

Source facts

Action
On 11/03/2020, a recall notice was emailed. Their customer was asked to do the following: 1 - Identify and quarantine all affected devices present in your establishment. 2 - Circulate this safety notice to all concerned. 3 - Complete return the customer response form . 4 - In the event that you have distributed the products concerned to another organization yourself, circulate this notice to them and then return the response forms to us once completed. 5 - Return the devices to us (or contact us to organize the removal) by attaching a copy of the attached customer response form. For any additional questions regarding this recall, please contact Anne-Ccile Grandremy, Quality and Regulatory Affairs Manager on 06 59 08 60 56 or by email acgrandremy@neosteo.com.
Initiated
2020-11-03
Root cause
Process control
Terminated
2021-12-06
Distribution
US Nationwide distribution in the state of New Jersey.
Recall number
Z-2236-2021
Recall status
Terminated
CDRH record ID
188354
FDA record created
2021-08-09
Recall event number
88307

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗