Skip to content

Public FDA record

Z-2236-2023: Trexo Device

Recall record Z-2236-2023 for Trexo Robotics Holdings Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2236-2023
Company in FDA record
Trexo Robotics Holdings Inc.
FDA classification date
Jul 25, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Electrical issues that could potentially render the device non-functional, give off electromagnetic influence to surrounding equipment, and risk of burns when in use.

Source facts

Action
The firm distributed recall notification to customers by email on 5/31/2023. This notification asks that customers not expose the device to rain or spilled liquids, not store the device outdoors, wear pants to cover exposed skin to minimize the effect of hot surfaces, and if experiencing any electrical interference or difficulty operating another device in the vicinity of a Trexo device move away and and/or turn off the Trexo Device. Affected devices will be subject to a correction which will be carried out by technician; technician appointments will be scheduled once customers reply to the recall notification. Customers with any questions or concerns may contact the firm at correction@trexorobotics.com. Customers are asked to return the provided response form.
Initiated
2023-05-31
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of the Argentina, Australia, Brazil, Canada, China, Croatia, Czech Republic, England, Germany, India, Kuwait, Mexico, Poland, Qatar, Republic of Cyprus, Serbia, Singapore, Slovakia, Spain, United Arab Emirates, and Yemen.
Recall number
Z-2236-2023
Recall status
Open, Classified
CDRH record ID
201662
FDA classified
2023-07-25
Recall event number
92563

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗