Open original FDA record ↗Recall record
- FDA identifier
- Z-2236-2024
- Company in FDA record
- Exactech, Inc.
- FDA classification date
- Jun 28, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.
Source facts
- Action
- Exactech issued an UREGNT VOLUNTARY MEDICAL DEVICE RECALL notice to its sales representatives on 04/26/2024 via email. The notice explained the issue, clinical impact (risk), and requested the following: Actions to be Taken: " Review this notification thoroughly. " Immediately discontinue use and quarantine any affected product. " Distribute this Recall Letter to all the surgeons who have implanted the affected devices. " Send all affected product back to Exactech as outlined in the attached Recall Confirmation Form .
- Initiated
- 2024-04-26
- Root cause
- Other
- Distribution
- Worldwide - US distribution in Puerto Rico and the countries of Great Britain, Japan.
- Recall number
- Z-2236-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207612
- FDA classified
- 2024-06-28
- Recall event number
- 94410