Open original FDA record ↗Recall record
- FDA identifier
- Z-2237-2010
- Company in FDA record
- Neuisys, LLC
- FDA classification date
- Aug 19, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The reference lines for image generated for a Surview scan may appear in the incorrect position in the "FILM" display mode.
Source facts
- Action
- On 07/21/2008, NEUISYS Imaging Systems Solutions distributor in Greensboro, NC began sending the URGENT DEVICE CORRECTION letter on behalf of Philips and NeuSoft Medical Systems (PNMS), Co., Ltd in China. The letter informs the consignees that there is potential safety problem in the NeuViz Dual series CT scanners due to the malfunction of "Incorrect Reference Line in the Film". When a "Head first, Couch out" or a "Feet first, Couch in" Surview scan is started then paused midstream, the image generated is what the doctor expects, but the reference lines appear in the incorrect position in "FILM" display mode. The acquired images accurately represent the exact positions planned on the original view. The incorrect/unmatched reference line in film may result in incorrect diagnosis. The consignees are advised not to stop or pause during Surview acquisition while PNMS in China will provide the updated software patch to all affected systems free of charge. The Field Change Order #FCOP-08-CT-NMS-008 SP4 will be implemented by NEUISYS Imaging Systems Solutions in Greensboro, NC., who is a distributor for PNMS. Consignees are advised to e-mail Service Support Department of PNMS at helpdesk@pnms.neusoft.com regarding the recall.
- Initiated
- 2008-07-21
- Root cause
- Software design
- Terminated
- 2010-08-23
- Distribution
- Nationwide Distribution
- Recall number
- Z-2237-2010
- Recall status
- Terminated
- CDRH record ID
- 72515
- FDA classified
- 2010-08-19
- Recall event number
- 49012