Skip to content

Public FDA record

Z-2237-2012: i-STAT cTnI cartridges Product Usage: The i-STAT cTnl test is an in vitro diagnostic test for the q

Recall record Z-2237-2012 for Abbott Point Of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2237-2012
Company in FDA record
Abbott Point Of Care Inc.
FDA classification date
Aug 22, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Abbott Point of Care has determined that i-Stat CTNL cartridges have the potential to exhibit incorrectly elevated cardiac troponin I results. This recall was expanded on July 19 2012 to include additional information regarding user documentation.

Source facts

Action
Abbott Point of Care (APOC) sent an Urgent Recall Notice letters, via Fed Ex on March 07, 2012 to all affected customers. APOC sent an important product information update on July 19, 2012 to all of their affected customers. The letter identified the affected product, problem and actions to be taken. Customers are asked to review the information included as it may pertain to their cartridges. Customers were instructed that if the affected product have been forwarded to another facility, provide a copy of this letter to them. For additional information, contact APOC at 800-366-8020 option 1 or your Abbott Point of Care representative.
Initiated
2012-03-07
Root cause
Device Design
Terminated
2013-02-19
Distribution
Worldwide Distribution - US Nationwide, International
Recall number
Z-2237-2012
Recall status
Terminated
CDRH record ID
110437
FDA classified
2012-08-22
Recall event number
62410

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.