Open original FDA record ↗Recall record
- FDA identifier
- Z-2237-2012
- Company in FDA record
- Abbott Point Of Care Inc.
- FDA classification date
- Aug 22, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Abbott Point of Care has determined that i-Stat CTNL cartridges have the potential to exhibit incorrectly elevated cardiac troponin I results. This recall was expanded on July 19 2012 to include additional information regarding user documentation.
Source facts
- Action
- Abbott Point of Care (APOC) sent an Urgent Recall Notice letters, via Fed Ex on March 07, 2012 to all affected customers. APOC sent an important product information update on July 19, 2012 to all of their affected customers. The letter identified the affected product, problem and actions to be taken. Customers are asked to review the information included as it may pertain to their cartridges. Customers were instructed that if the affected product have been forwarded to another facility, provide a copy of this letter to them. For additional information, contact APOC at 800-366-8020 option 1 or your Abbott Point of Care representative.
- Initiated
- 2012-03-07
- Root cause
- Device Design
- Terminated
- 2013-02-19
- Distribution
- Worldwide Distribution - US Nationwide, International
- Recall number
- Z-2237-2012
- Recall status
- Terminated
- CDRH record ID
- 110437
- FDA classified
- 2012-08-22
- Recall event number
- 62410