Open original FDA record ↗Recall record
- FDA identifier
- Z-2237-2023
- Company in FDA record
- Beckman Coulter, Inc.
- FDA classification date
- Jul 26, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
Source facts
- Action
- A customer letter titled, Urgent Medical Device Recall, dated 06/07/2023, was issued to the impacted consignee via first-class mail and email. The consignee is instructed to discard the product and request a replacement. The consignee is asked to complete the response form and submit to the recalling firm.
- Initiated
- 2023-06-20
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.
- Recall number
- Z-2237-2023
- Recall status
- Open, Classified
- CDRH record ID
- 201757
- FDA classified
- 2023-07-26
- Recall event number
- 92576