Z-2237-2025: BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD

Recall record Z-2237-2025 for Becton, Dickinson and Company, BD Bio Sciences. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2237-2025
Company in FDA record
Becton, Dickinson and Company, BD Bio Sciences
FDA classification date
Aug 1, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.

Source facts

Action
On July 8, 2025, BD Biosciences issued a "Urgent: Medical Device Recall Correction" via mail. BD ask consignees to take the following actions: 1. Ensure the contents of this product notification are read and understood. 2. Share and post this customer letter with the users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 3. It is recommended that this notice be posted where these products are stored. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted product, so that BD may acknowledge your receipt of this notification. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.
Initiated
2025-07-08
Root cause
Labeling design
Distribution
Worldwide - US Nationwide distribution in the states of Tennessee, New York, Massachusetts and the countries of Argentina, Belgium, Botswana, Brazil, Canada, France, Germany, Lithuania, Peru, Philippines, United Kingdom.
Recall number
Z-2237-2025
Recall status
Open, Classified
CDRH record ID
214647
FDA classified
2025-08-01
Recall event number
97141

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗