Skip to content

Public FDA record

Z-2238-2019: Quantum Pump Console

Recall record Z-2238-2019 for Spectrum Medical Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2238-2019
Company in FDA record
Spectrum Medical Ltd.
FDA classification date
Aug 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.

Source facts

Action
Spectrum Medical notified customers on about 05/29/2019 via "URGENT FIELD SAFETY CORRECTION" letter. Customers were informed that Spectrum Medical is in the process of installing the most recent software update v6.2.3.16 to the Quantum Perfusion Systems, (Quantum Pump Console). This software update corrects an error that could potentially lead to power loss. All Quantum Perfusion Systems are in scope of this issue. Customers were instructed to be aware that user intervention may be necessary with continued use before the software update, assure that all users receive notice of this issue, confirm receipt of the Urgent Medical Device Correction by returning the customer response form, and a Spectrum Medical service team will contact each customer and arrange a time and date to update the systems with the most recent software update. Additionally, customers were instructed to notify all relevant personnel and to customers if the devices have been further distributed.
Initiated
2019-05-29
Root cause
Software design
Terminated
2020-08-21
Distribution
US distribution to NY, OH, VA, WA. International distribution to the Netherlands.
Recall number
Z-2238-2019
Recall status
Terminated
CDRH record ID
174881
FDA classified
2019-08-23
Recall event number
83352

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗