Open original FDA record ↗Recall record
- FDA identifier
- Z-2239-2008
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Sep 19, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When running the Automated Dilution Protocol with the assay files "CMV IgG" or "Toxo IgG" on the ARCHITECT i1000SR system, a software error [9010 (Pipettor Robotics) preventing (Process Path) from operating] is generated and the instrument stops running. No results are generated.
Source facts
- Action
- Abbott affiliates were e-mailed copies of the recall letter Urgent Field Safety Notice Product Correction on 6/16/08 for translation where needed and dissemination to their ARCHITECT accounts who have the affected assay CD-ROMs. Each Abbott international organization will be responsible for contacting customers and determining the method of effectiveness checks based on their country requirements. Abbott sent Product Correction, Immediate Action Required letters dated June 16, 2008 to all active non-U.S. ARCHITECT customers who have the affected assay CD-ROMs. The end users were informed when running the Automated Dilution Protocol of the ARCHITECT CMV IgG and Toxo IgG assays on the ARCHITECT i1000SR System, software error code [9010 (Pipettor Robotics) preventing (Process Path) from operating] is generated and the instrument stops. The accounts were advised to not use the Automated Dilution Protocol of either the ARCHITECT CMV IgG and Toxo IgG assays on the ARCHITECT i1000SR System and report results for samples requiring dilution as ">250.0 AU/mL" for CMV IgG and ">200.0 IU/mL" for Toxo IgG until a corrected version of the ARCHITECT i1000SR System Assay CD-ROM is available in late September 2008. The accounts were requested to keep a copy of the letter on file. The accounts were requested to complete the customer reply form, indicating that the letter was received, understood and disseminated to the laboratory manager/supervisor responsible for Architect testing, and fax it to Abbott by 7/1/08 at 1-800-777-0051. If you have question, contact your local area Customer Support.
- Initiated
- 2008-06-16
- Root cause
- Software design
- Terminated
- 2010-03-18
- Distribution
- Austria, Belgium, Czech Republic, Denmark, France, Germany, Iceland, India, Italy, Latvia, Lithuania, Netherlands, Poland, Portugal, Sweden, Saudi Arabia, Spain, United Arab Emirates
- Recall number
- Z-2239-2008
- Recall status
- Terminated
- CDRH record ID
- 71929
- FDA classified
- 2008-09-19
- Recall event number
- 48701