Open original FDA record ↗Recall record
- FDA identifier
- Z-2239-2010
- Company in FDA record
- Starion Instruments
- FDA classification date
- Aug 19, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Starion Instruments Thermal Ligating Shears (STLS) 35C and the Starion Thermal Ligating Shears (STLS) 23C may turn on and remain on without the finger button being depressed. The devices may also work intermittently, incorrectly indicating the device is 'on".
Source facts
- Action
- Starion distributed a "STLS Device Recall - Important Medical Device Information" letter dated March 9, 2010 to consignees. The letter identified the affected model and lot numbers. Consignees were asked to complete a inventory status form and fax to Starion. the letter included a shipping labe with a pre-assigned return authorization number. Starion will ship replacement after receipt of the inventory status form and returned product..
- Initiated
- 2010-03-09
- Root cause
- Device Design
- Terminated
- 2010-11-27
- Distribution
- Worldwide Distribution: USA states of CA, OH, MI, KS and MO and countries of Italy, China, Kuwait, and Germany.
- Recall number
- Z-2239-2010
- Recall status
- Terminated
- CDRH record ID
- 89744
- FDA classified
- 2010-08-19
- Recall event number
- 55043