Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.
Source facts
Action
Philips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated August 8, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Field Service Engineers will be implementing a Field Change Order (FCO) to replace the batteries of the affected UPS units. Contact the Philips Healthcare Customer Care Solutions Center at 1-800-722-9377 for questions regarding this notice.
.
Initiated
2012-08-08
Root cause
Material/Component Contamination
Terminated
2014-04-28
Distribution
Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA.
Recall number
Z-2239-2012
Recall status
Terminated
CDRH record ID
111947
FDA classified
2012-08-22
Recall event number
62892