Z-2239-2024: Avalon Ultrasound Transducer. Directs low-energy ultrasound beam toward the fetal heart and externally mo

Recall record Z-2239-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2239-2024
Company in FDA record
Philips North America Llc
FDA classification date
Jun 28, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for inaccurate fetal heart rate measurements when monitoring multiples.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notice dated 5/22/2024 was mailed to consignees. Consignees are instructed to identify what software their Avalon wired transducers. Avalon wired transducers with software revision L.01.04 must not be used when monitoring multiples and can only be safely used when monitoring the fetal heart rate of singletons. Avalon wired transducers that do not possess the affected software revision can be used when monitoring twins or triplets. An alternative to the Avalon wired Transducer is using the wireless Avalon CL Ultrasound transducer. Consignees are to complete and return the provided response form at recall.responsephilips.com or by fax to 877-499-7223. The provided recall notification should be shared with all clinical staff and kept with the documentation of the Avalon ultrasound transducer. Consignees with questions should contact their local Philips representative or call 1-800-722-9377.
Initiated
2024-05-22
Root cause
Component change control
Distribution
Worldwide distribution - US Nationwide and the countries of AD, AE, AR, AT, AU, BA, BE, BG, BH, BO, BR, CH, CL, CO, CY, CZ, DE, DZ, EG, ES, FI, FR, GB, GI, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JP, KE, KW, LB, LK, LS, LT, LU, LV, LY, MA, ME, MT, MU, MY, NA, NI, NL, NZ, OM, PA, PH, PK, PL, PR, PS, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, VN, ZA.
Recall number
Z-2239-2024
Recall status
Open, Classified
CDRH record ID
208256
FDA classified
2024-06-28
Recall event number
94774

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗