Open original FDA record ↗Recall record
- FDA identifier
- Z-2240-2016
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Jul 19, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The screw may fracture at the junction of the shaft and the head during implantation. The patient may retain a foreign body if the fractured screw cannot be removed. Soft tissue irritation will not occur as the retained portion of the screw is below the surface of the plate. A delay in surgery less than 30 minutes may occur.
Source facts
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall letter dated June 10, 2016, to all affected customers via courier. Distributors and hospitals were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-6725, Monday through Friday, 8 a.m. to 5 p.m.
- Initiated
- 2016-06-10
- Root cause
- Device Design
- Terminated
- 2017-07-20
- Distribution
- Worldwide Distribution - US (Nationwide) Internationally to Australia, The Netherlands, and Japan
- Recall number
- Z-2240-2016
- Recall status
- Terminated
- CDRH record ID
- 147936
- FDA classified
- 2016-07-19
- Recall event number
- 74576