Skip to content

Public FDA record

Z-2240-2021: Covidien Sonicision Reusable Generator-converts electrical power from the battery pack to ultrasonic ener

Recall record Z-2240-2021 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2240-2021
Company in FDA record
Covidien Llc
FDA record creation date
Aug 10, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for a manufacturing assembly error-may result in a non-functional audio indicator of device activation or a non-functional device, may cause unintended thermal burn, hemorrhage, or tissue damage and a delay of treatment while an alternate device is located

Source facts

Action
Covidien notified consignees by Urgent Medical Device Recall letter delivered via Federal Express and/or certified mail beginning June 23, 2021. Letter states reason for recall, health risk and action to take: 1. Please immediately quarantine and discontinue use of the affected item code with associated lot and serial numbers listed on Attachment A. 2. Please return affected product as indicated below. All products from the affected item code and associated lot and serial numbers must be returned. 3. If you have distributed the Covidien Sonicision" Reusable Generators listed on Attachment A, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. We request that you contact Medtronic if you experienced a quality problem or adverse event. Email Medtronic Post Market Vigilance at: rs.assurancequality@medtronic.com Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Initiated
2021-06-23
Root cause
Under Investigation by firm
Terminated
2025-01-28
Distribution
US Nationwide Distribution: AL, AZ, CT, GA, IN, MA, MD, MN, NY, OH, TX, VA and WA.
Recall number
Z-2240-2021
Recall status
Terminated
CDRH record ID
188264
FDA record created
2021-08-10
Recall event number
88271

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗