Open original FDA record ↗Recall record
- FDA identifier
- Z-2241-2010
- Company in FDA record
- Dental Equipment LLC
- FDA classification date
- Aug 20, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The magnetic headrest on certain dental chairs may affect the function/programming of some implantable pacemakers or defibrillators if the implanted device is programmed to respond to a magnet.
Source facts
- Action
- DCI Equipment sent an "Urgent Field Safety Alert" letter dated March 15, 2010 to consignees advising them of the potential problem. The letter instructed that patients with implantable pacemakers or defibrillators should avoid dental chairs with the magnetic headrest. A warning label was sent with the letter for insertion into the product Use and Care manual for operator warning. Consignees were asked to complete a Field Safety Alert Acknowledgement/Return Form and fax to the firm. Consignees were requested to notify customers.Consignees may contact the firm at 503 537-3602.
- Initiated
- 2010-03-23
- Root cause
- Other
- Terminated
- 2012-01-04
- Distribution
- Worldwide distribution in the USA and Canada.
- Recall number
- Z-2241-2010
- Recall status
- Terminated
- CDRH record ID
- 90151
- FDA classified
- 2010-08-20
- Recall event number
- 55153