Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image.
Source facts
Action
Neusoft Medical Systems sent an Urgent Device Correction Notification letter dated February 9, 2011, via certified mail to their distributor NEUISYS. The letter identified the product, the problem, and the action to be taken by the customer.
Customers were advised to discontinue use of combined images. Instead, use MPR Batch for diagnostic reviews until the planned upgrade is provided to the product.
Neusoft Medical System and Philips Medical System released the mandatory Field Change Order (FCO) by March 31, 2011. All nine affected systems will be updated with aforementioned kits in six months after the release.
Consignees were instructed to contact Service Support Department by e-mail to nms-service@neusoft.com
Initiated
2011-02-09
Root cause
Other
Terminated
2011-08-24
Distribution
Nationwide Distribution including CA, GA, LA, NC, OH, TN and TX.
Recall number
Z-2241-2011
Recall status
Terminated
CDRH record ID
99626
FDA classified
2011-05-23
Recall event number
58582