GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
Source facts
Action
No customer communication took place. On April 26, 2010, GE began opening service cases for field engineers to begin replacing the snubber boards. This action has been completed as of October 29, 2010.
Initiated
2010-04-26
Root cause
Component change control
Terminated
2012-08-22
Distribution
Worldwide and Nationwide (USA) distribution.
Recall number
Z-2241-2012
Recall status
Terminated
CDRH record ID
111516
FDA classified
2012-08-22
Recall event number
62729