Open original FDA record ↗Recall record
- FDA identifier
- Z-2241-2024
- Company in FDA record
- Siemens AG/Siemens Healthcare GmbH
- FDA classification date
- Jul 2, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
Source facts
- Action
- Inform customers of the temporary workaround to mitigate the issue: Take any action which will lead to a change of the prefiltration prior to the first x-ray release after a patient registration. This could be the first acquisition, the change of the organ program or change to water values due to steeper angulation. This will help to avoid the occurrence of this issue. Siemens Medical Solutions will contact customers to arrange a date to distribute a software update to permanently resolve the defect.
- Initiated
- 2024-06-26
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
- Recall number
- Z-2241-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208750
- FDA classified
- 2024-07-02
- Recall event number
- 94892