Open original FDA record ↗Recall record
- FDA identifier
- Z-2242-2014
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Aug 15, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
Source facts
- Action
- Urgent Medical Device Recall Notices dated July 15, 2014 were mailed with verification of delivery via FedEx. Response forms were provided to all consignees in order for Biomet to track returns. The letters informed the consignees to return the devices to Biomet and conduct a sub-recall sending the recall letter to all customers.
- Initiated
- 2014-07-01
- Root cause
- Process change control
- Terminated
- 2015-10-30
- Distribution
- California, IA, FL, MD, UT, SC, AL, IL, CO, MA, KS, GA, WA, NV, AR, NH, OK,Pennsylvania, New York, Texas, Virginia, New Jersey, North Carolina, Kentucky, Michigan, Ohio, Wisconsin, Louisiana, Arizona, Indiana, South Dakota, Missouri & Montana. In addition to Korea, B.V., Japan, Puerto Rico, UK, Chile, New Zealand, Australia, & Canada
- Recall number
- Z-2242-2014
- Recall status
- Terminated
- CDRH record ID
- 129028
- FDA classified
- 2014-08-15
- Recall event number
- 68931