Z-2243-2009: TomoTherapy Hi-Art System, Version 3.X ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The

Recall record Z-2243-2009 for TomoTherapy Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2243-2009
Company in FDA record
TomoTherapy Incorporated
FDA classification date
Sep 21, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

TomoTherapy received one customer report indicating that after starting a patient procedure from the Status Console, the radiation on light and the audible indicator on the Status Console came on as expected, however the Operator Station (OS) did not display the dose rate and the procedure time, or the couch position updates. Within a few minutes, the customer pressed the Stop button on the Sta

Source facts

Action
Consignees were sent a TomoTherapy "Urgent Medical Device Correction Field Safety Notice" letter dated August 10, 2009. The letter addressed the Issue, Product Affected, Investigation, Recommended Actions and Resolution. Customers were recommended to perform the following actions (in the event that the dose rate, procedure timer, and couch position updates are not present on the OS, after pressing Start on the status console and the radiation on light and audible indicators are active): 1) Immediately press the Stop button on the Status Console and 2) contact your authorized TomoTherapy Service Representative for further direction. The affected issue will be corrected in the next software version, which is scheduled for release in early 2010. The TomoTherapy Customer Contact Center can be reached 24 hours a day, 7 days a week, by e-mail at support@tomotherapy.com or by phone at 1-866-368-4807 (North America).
Initiated
2009-08-10
Root cause
Software design
Terminated
2011-12-17
Distribution
Worldwide Distribution -- US (states of AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, HI, IL IN, IA, KS, KY, LA, MD, MA, MI, MN,MO, NE,NV,NJ, NM, NY, OH, OK OR, PA, PR, SC, SC, TN, TX, UT, VA, WA, WI and WY) and Belgium Canada Switzerland China Germany Spain France England Hong Kong India Italy Japan Korea Netherlands Poland Saudi Arabia Sweden Singapore Turkey and Taiwan.
Recall number
Z-2243-2009
Recall status
Terminated
CDRH record ID
83902
FDA classified
2009-09-21
Recall event number
52729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗