Skip to content

Public FDA record

Z-2243-2012: Sunquest Encompass The reporting site is using the Encompass software in an acute patient care setting

Recall record Z-2243-2012 for Sunquest Information Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2243-2012
Company in FDA record
Sunquest Information Systems, Inc.
FDA classification date
Aug 22, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class III

Reason stated in the source

Sunquest is recalling the Sunquest Encompass versions 2.4 and later because the specimen collect time is displaying and printing one hour later than the actual collect time only when Daylight Saving Time is active.

Source facts

Action
The firm, Sunquest, sent a "Urgent-Product Safety Notice: PSN-09-01" letter dated April 13, 2009 all their customers who purchase the Sunquest Encompass versions 2.4 and later. The letter informed the customers of the problem identified and the action to be taken. Customers were instructed to request software correction and to create a service request via client support webpage at www.sunquestinfo.com/supportweb. Then choose Submit a Service Request. Select Issue Type and select Software Request. When Software Request is selected, the Work Queue field automatically fills in with "Software Request." Promptly advise appropriate personnel. Customers with questions can contact Client Support and Services at (877) 239-6337 for US and Canadian sites and at +44(0)0800-960-972 for international sites.
Initiated
2009-04-13
Root cause
Software design
Terminated
2012-08-22
Distribution
Worldwide distribution: USA (nationwide) and country of: Bahamas.
Recall number
Z-2243-2012
Recall status
Terminated
CDRH record ID
111640
FDA classified
2012-08-22
Recall event number
62751

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.