Incorrect transponder data was written on the RFID tags
triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery
Source facts
Action
Smith & Nephew Issued Urgent Field Safety Notice (FSN) - MEDICAL DEVICE RECALL
WOM reference no.: 2024-0001 via email on 5/24/23. Letter states reason fo recall, health risk and action to take;
lnform individuals within your organization who need to be aware of this recall.
- Check all stock areas and/or operating room storage to determine if any tube sets with lol numbers from Annex 'l are at your facility,
- Dispose of all affected tube sets properly. Confirm this on the customer response form. Return the form to fieldactions@smith-nephew.com.
Please send any questions to fieldactions@smith-neDhew.com
Initiated
2024-05-24
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-2243-2024
Recall status
Open, Classified
CDRH record ID
208168
FDA classified
2024-07-01
Recall event number
94749
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.