Skip to content

Public FDA record

Z-2244-2011: BacT/ALERT - FA Reagent Bottle The BacT/ALERT Microbial Detection System is used to determine if microorg

Recall record Z-2244-2011 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2244-2011
Company in FDA record
bioMerieux, Inc.
FDA classification date
May 24, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The charcoal suspension in a portion of the lot does not meet the intended density.

Source facts

Action
BioMerieux sent an Urgent Product Removal Notice dated April 13, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Evaluate positive results adhering to their institution's protocols for determining false versus true positives. Negative bottles should be checked by smear and/or subculture at some point prior to discarding as negative. Any remaining inventory of BacT/ALERT FA culture bottle lot 1027481 should be destroyed in accordance with their facility's policies and procedures. An Acknowledgement and Product Replacement Form was included to be completed and returned to bioMerieux or credit or replacement of unused bottles.
Initiated
2011-04-13
Root cause
Other
Terminated
2012-04-03
Distribution
Worldwide Distribution - USA (nationwide) and the countries of China, Tokyo, South Korea, Canada, Sweden, Austria, Poland, and Belgium
Recall number
Z-2244-2011
Recall status
Terminated
CDRH record ID
99580
FDA classified
2011-05-24
Recall event number
58550

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗