A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.
Source facts
Action
Medtronic Navigation issued an "Urgent Field Safety Notice" to consignees dated 7/11/12. The problem was described and recommended actions were provided. For each affected O-arm Imaging System, a Medtronic service representative will visit the site and replace the 500VDC relay on the system and test the O-arm Imaging System.
Initiated
2012-07-11
Root cause
Component change control
Terminated
2016-06-06
Distribution
Worldwide Distribution - USA, including the states of AL, CO, FL, IA, MI, IA, NC, NE, NY and the countries of South Africa, Italy, and Spain.
Recall number
Z-2244-2012
Recall status
Terminated
CDRH record ID
111068
FDA classified
2012-08-22
Recall event number
62571