Skip to content

Public FDA record

Z-2244-2021: Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to

Recall record Z-2244-2021 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2244-2021
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Aug 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

Source facts

Action
Cardinal Health issued Urgent Medical Device Correction letter and Acknowledgement Form on June 16, 2021 to the direct consignees. Letter states reason for recall, health risk and action to take: Actions Required: 1) REVIEW the IFU for the Safety Scalpel N11 (vendor part number 73-1811). 2) COMMUNICATE the IFU to all personnel that utilize the Argyle" UVC Insertion Tray (SKU 43201) containing Safety Scalpel N11 (vendor part number 73-1811) 3) NOTIFY any customers to whom you may have distributed/forwarded affected product or will send the product about this product correction notice and share a copy of this notice and the attached IFU. 4) POST a copy of this notification in the location where the product is stored. 5) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB-FieldCorrectiveAction@cardinalhealth.com whether or not you have affected product. Available Assistance: Please contact the Customer Service group for any questions or to arrange for credit and return of any product: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All other customers 888-444-5440 For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Initiated
2021-06-16
Root cause
Labeling Change Control
Terminated
2023-01-04
Distribution
US Nationwide distribution.
Recall number
Z-2244-2021
Recall status
Terminated
CDRH record ID
188253
FDA classified
2021-08-13
Recall event number
88266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗