Z-2244-2023: Ventralight ST w EchoPS 4"x6" Ellipse- A low profile, bioresorbable, coated, permanent mesh, with a pre-a

Recall record Z-2244-2023 for Davol, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2244-2023
Company in FDA record
Davol, Inc.
FDA classification date
Jul 27, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

Source facts

Action
BD issued Urgent Medical Device Product Recall letter on 4/25/23 via FedEx 2-day and email. Letter states reason for recall, health risk and action to take: 1. Please immediately discontinue use of the affected product. Check all inventory locations within your facility and destroy all affected product in accordance with your facility s process. 2. Share this notification with all users of the affected product within your facility network to ensure they are also aware of this product recall. 3. If you purchased the affected product from a distributor, contact your distributor directly for further instructions and credit resolution. 4. Complete the attached Customer Response Form and return it to the BD contact noted on the form whether or not you have any of the affected product so that BD may acknowledge your receipt of this notification per FDA requirements. 5. Indicate on the response form the quantity of affected product identified at your facility and confirm that this product inventory was destroyed. 6. Please contact your BD representative if you require assistance with this process, or you may also contact BD. 1. BD will issue replacements for destroyed product to customers that purchased product directly from BD following receipt of the completed Customer Response Form. 2. BD is implementing actions to prevent recurrence of this product issue. Contact Information: If you require further assistance please contact: BD Contact Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD- LIFE (1-844-823-5433) Say Recall when prompted M-F 8am - 5pm CT Email: productcomplaints@bd.com Recall questions, Product Complaints, Technical Questions
Initiated
2023-04-25
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.
Recall number
Z-2244-2023
Recall status
Open, Classified
CDRH record ID
199946
FDA classified
2023-07-27
Recall event number
92213

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗