Z-2244-2024: EVOLUTION MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cemen

Recall record Z-2244-2024 for MicroPort Orthopedics Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2244-2024
Company in FDA record
MicroPort Orthopedics Inc.
FDA classification date
Jul 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3 Evolution MP Non-Porous Femoral Left in the packaging.

Source facts

Action
The recalling firm issued the FSN, dated 05/02/2024, via email. The required actions were to: Immediately check inventory and quarantine all subject product(s); COMPLETE AND RETURN the attached FSCA Acknowledgement; Inform MicroPort Orthopedics of any adverse event immediately; Return any affected product to: MicroPort Orthopedics Inc. 1481 Gulf Stream, Arlington, TN 38002. The consignee is to mark all return shipping boxes with RECALL on several sides for better identification and processing.
Initiated
2024-05-03
Root cause
Under Investigation by firm
Distribution
International distribution in the country of China. There is no field inventory in the United States for either lot.
Recall number
Z-2244-2024
Recall status
Open, Classified
CDRH record ID
208115
FDA classified
2024-07-01
Recall event number
94729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗