Skip to content

Public FDA record

Z-2245-2021: FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No:

Recall record Z-2245-2021 for BioFire Diagnostics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2245-2021
Company in FDA record
BioFire Diagnostics, LLC
FDA record creation date
Aug 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.

Source facts

Action
On 07/20/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customer informing them that an issue has been identified in the manufacturing process which might contribute to elevated rates of false negative results while using the BCID Panel. Actions to be taken by customers: 1) Immediately examine their inventory for product identified in this recall, 2) Discontinue use and discard any remaining product in their possession that is subject to this recall. The Recalling Firm will replace the product at no charge in accordance with its standard limited warranty. 3) Complete the accompanying Field Action Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to the Recalling Firm so that it may acknowledge their receipt of this notification. 4) If they may have further distributed the affected product, identify any recipients and notify them at once of this product recall. For product replacement, as well as any other questions or concerns, contact customer support department at Support@biofiredx.com or via telephone at 1-800-735-6544 and selecting option 5 for Product Technical Support.
Initiated
2021-07-20
Root cause
Process control
Terminated
2022-12-16
Distribution
U.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None
Recall number
Z-2245-2021
Recall status
Terminated
CDRH record ID
188591
FDA record created
2021-08-12
Recall event number
88410

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗