Skip to content

Public FDA record

Z-2246-2024: First Source Vision M. Portable cart and platform that supports an X-ray system and laptop computer.

Recall record Z-2246-2024 for First Source Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2246-2024
Company in FDA record
First Source Inc
FDA classification date
Jul 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the bolt that secures the positioning arm to the gas spring to come loose and fall onto patient.

Source facts

Action
Consignees were emailed an URGENT: MEDICAL DEVICE RECALL notification, dated 5/29/24. The notice instructs consignees to follow the recommended preventive maintenance schedule and visually check the device before use to identify if the affected bolt is coming loose. The recalling firm is currently testing a spring clip which will keep the bolt from falling out and will send this part to customers at no charge once approved. Consignees are asked to return the provided response form. Consignees with any questions can contact the firm by email at support@fsimed.com or by phone at 1-800-349-5980 (option 2) from 9:00 AM to 5:00 PM EST from Monday through Friday.
Initiated
2024-05-09
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, IL, IN, LA, MA, MD, MI, MO, MS, NC, NY, OK, PA, TN, TX, & VA. The countries of Korea & Serbia.
Recall number
Z-2246-2024
Recall status
Open, Classified
CDRH record ID
208161
FDA classified
2024-07-01
Recall event number
94744

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗