Z-2247-2017: KYPHON(R) Directional Bone Filler Device, Size: 3, REF F04C, Rx Only, STERILE R.

Recall record Z-2247-2017 for Medtronic Sofamor Danek USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2247-2017
Company in FDA record
Medtronic Sofamor Danek USA Inc
FDA record creation date
Jun 3, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The directional arrow at the proximal end of the instrument may not correctly align with the cutout opening on the distal end of the instrument. Using affected bone void filler may result in the injected cement being placed in a direction unintended by the surgeon.

Source facts

Action
Medtronic distributed a voluntary removal notice (Attachment I) via FedEx courier service to impacted U.S. consignees on 17 May 2017. Communications to impacted consignees outside the United States are currently ongoing per each geographies local procedures. The consignees were directed to do the following: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via fax to 763-367-8134, or via e-mail to rs.sdmriskmgt@medtronic.com. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product.
Initiated
2017-05-17
Root cause
Nonconforming Material/Component
Terminated
2019-06-18
Distribution
worldwide
Recall number
Z-2247-2017
Recall status
Terminated
CDRH record ID
155810
FDA record created
2017-06-03
Recall event number
77352

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗