Open original FDA record ↗Recall record
- FDA identifier
- Z-2247-2017
- Company in FDA record
- Medtronic Sofamor Danek USA Inc
- FDA record creation date
- Jun 3, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The directional arrow at the proximal end of the instrument may not correctly align with the cutout opening on the distal end of the instrument. Using affected bone void filler may result in the injected cement being placed in a direction unintended by the surgeon.
Source facts
- Action
- Medtronic distributed a voluntary removal notice (Attachment I) via FedEx courier service to impacted U.S. consignees on 17 May 2017. Communications to impacted consignees outside the United States are currently ongoing per each geographies local procedures. The consignees were directed to do the following: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via fax to 763-367-8134, or via e-mail to rs.sdmriskmgt@medtronic.com. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product.
- Initiated
- 2017-05-17
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-06-18
- Distribution
- worldwide
- Recall number
- Z-2247-2017
- Recall status
- Terminated
- CDRH record ID
- 155810
- FDA record created
- 2017-06-03
- Recall event number
- 77352