Skip to content

Public FDA record

Z-2247-2023: VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Recall record Z-2247-2023 for Johnson & Johnson Surgical Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2247-2023
Company in FDA record
Johnson & Johnson Surgical Vision, Inc.
FDA classification date
Jul 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Source facts

Action
On 5/30/23, recall notices were mailed to customers who were asked to do the following: 1) Discontinue use and return affected devices to the recalling firm. 2) Share this notice with anyone within your organization that needs to be informed and to any organization where the potentially affected products have been transferred. 3) Complete and return the customer reply form to RegCompliOne@its.jnj.com In addition, distributors were asked to distribute the recall notice to their customers. Questions, adverse events, and complaints can be directed to the firm at 1-877-266-4543 Option 2.
Initiated
2023-05-30
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Korea¿(South), Kuwait, Lebanon, Malaysia, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.
Recall number
Z-2247-2023
Recall status
Open, Classified
CDRH record ID
201385
FDA classified
2023-07-28
Recall event number
92539

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗