Open original FDA record ↗Recall record
- FDA identifier
- Z-2248-2016
- Company in FDA record
- Steris Corporation
- FDA record creation date
- Jul 20, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
STERIS has identified that the pivot pin assembly located on the table top may become loose preventing the table from articulating properly.
Source facts
- Action
- Recall Notification Letters to the affected Consignees was sent on May 23, 2016 via FedEx with tracking numbers for delivery confirmation. STERIS Field Service Technicians will visit each domestic Customer location to perform the inspection and reinstallation of the pivot pin assemblies. For international direct accounts who have further distributed the surgical table, Steris asked to identify the Customers, and to notify them of this product field correction, and that not to operate the tables until the inspection has been done of the pivot pin assembly on all affected tables.
- Initiated
- 2016-05-23
- Root cause
- Other
- Terminated
- 2016-08-22
- Distribution
- AL;AK, AZ; AR; CA; CO; CT; DC; FL; GA; IL; IN; IA; NY; NJ; NC; OH; OK; OR; PA; PR; SC; TX; UT and VA. Foreign accounts: Canada; France; Korea; Panama; Qatar and Thailand.
- Recall number
- Z-2248-2016
- Recall status
- Terminated
- CDRH record ID
- 147015
- FDA record created
- 2016-07-20
- Recall event number
- 74264