The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.
Source facts
Action
Stryker Communications sent an Urgent Medical Device Recall letter to all affected customers. A Stryker Representative will contact their facility to schedule a service visit. to replace the adapter assembly to resolve the tolerance issue affected units. Customers with questions were instructed to Stryker Technical Support at 800-243-5135 or comm.techservice@stryker.com.
For questions regarding this recall call 972-834-8656.
Initiated
2017-04-21
Root cause
Component design/selection
Terminated
2017-09-20
Distribution
Worldwide Distribution - US and Canada
Recall number
Z-2248-2017
Recall status
Terminated
CDRH record ID
155570
FDA record created
2017-06-05
Recall event number
77261